Dr Jeff Domsic, Product Manager Charles River Laboratories, discusses ATP-bioluminescence testing aligned with new regulatory guidelines.
Over the past decade, the emergence of advanced therapy medicinal products (ATMPs) has fundamentally changed the approach to developing, manufacturing, administering, and treating diseases. These advanced modalities show significant promise and introduce various safety testing challenges. Industry and regulatory bodies have provided updated guidance regarding safe manufacturing and testing1, 2, 3.
ATP-bioluminescence technology has been validated and included in multiple regulatory approvals for routine release of a range of drug substances and products, including small molecule, biologics, vaccines, and others4. To fully utilise this rapid microbiological method (RMM), a key limitation of ATP-bioluminescence had to be addressed: the depletion of human cell line ATP from test samples. Charles River Laboratories has developed a novel methodology for removing eukaryotic cells while keeping the microbial cells intact. After processing the sample, an ATP-bioluminescence assay is run on a luminometer to detect microbial contamination in the cell-containing samples.
Charles River Laboratories published a case study describing this method, in which Jurkat T-cells were tested using ATP-bioluminescence5, 6. The data in Figure 1 confirmed that the T-cells tested did not interfere with or enhance the assay, thus demonstrating the technology’s feasibility.

ATP-bioluminescence method validation for cell-based testing
Following the guidance of Ph. Eur. 2.6.27 and USP <1071>, users are expected to perform validation studies to implement alternative microbiological methods such as ATP-bioluminescence. To help reduce the validation burden, Charles River Laboratories commissioned studies in accordance with USP <1223>“Validation of Alternative Microbiological Methods”, Ph. Eur. 5.1.6 “Alternative Methods for Control of Microbiological Quality” and PDA Technical Report No. 33, Revised 2013 “Evaluation, Validation and Implementation of Alternative and Rapid Microbiological Methods” that satisfied criteria required for validating a qualitative ATP-bioluminescence method7.
The company tested equivalency on leukapheresis samples6. Results of the four-day ATP method were compared to a 14-day USP <71> sterility test run side-by-side, with samples tested in parallel. The study referenced USP <1223> and used “decision equivalence” by acknowledging the binary nature of the test results in the form of presence (positive) or absence (negative)8. A sample size of 120 samples per method, divided equally among the organisms at the 1 CFU inoculum level, was chosen to statistically calculate if the ATP test is non-inferior to the 14-day USP <71> sterility test.

The results illustrated in Table 1 were used to calculate the lower one-sided confidence interval for the difference in detection probabilities. The acceptance criteria for a non-inferiority margin of Δ = 0.2 was recommended in USP <1223>. The calculated difference in detection probabilities of -0.0553 is within the non-inferiority limit, so the null hypothesis of inferiority can be rejected. Thus, the ATP method is statistically non-inferior to the USP <71> reference method.
CGT evaluated products proven compatible with the ATP method
Product types below were evaluated and confirmed to be compatible with ATP-bioluminescence.
Cultured Hematopoietic Stem and Progenitor Cell Therapies:
- Induced pluripotent stem cell (iPSC)-derived hematopoietic stem cells for regenerative medicine
- Autologous stem cell therapies for autoimmune diseases
- Allogeneic hematopoietic stem cell transplants for hereditary disorders and blood cancers
Viral Suspension Cell-Based Products:
- Lentiviral and retroviral vectors for gene therapy applications targeting genetic disorders
- Ex vivoviral vector-mediated gene editing tools, such as CRISPR/Cas9 systems for targeted gene correction
Lymphoblast Cell Lines for Research and Therapeutic Applications:
- Engineered T-cell therapies, including CAR-T cell therapies for oncological applications
- Immortalised B-cell lines for antibody production and research
Peripheral Blood Mononuclear Cells, both Fresh and Cryopreserved:
- Dendritic cell vaccines for cancer immunotherapy
- Natural killer (NK) cell therapies for infectious diseases and cancer
Cell Therapy for Glioblastoma using Implantable Devices:
- Oncolytic virus therapies targeting glioblastoma
- Stem cell-based regenerative therapies for neurological conditions
The future of atp-bioluminescence post-August 2025
Recent updates to USP have been added to provide additional guidance for selecting and qualifying RMMs for short-life products. In February 2025, USP published an update to USP <1071> “Rapid Microbiological Methods for the Detection of Contamination in Short-life Products – A Risk-based Approach” and introduced General Chapter <73> “ATP Bioluminescence-Based Microbiological Methods for the Detection of Contamination in Short-Life Products8.”
USP <73> guides risk-based testing for detecting microbial contamination in short-life and/or rapidly administered products and mandates that the primary validation of the method as described in <1071> must be available9. USP <1071> clearly states that if the RMM is not described in a USP general method chapter, then validation is required according to USP <1223>3. If the RMM is described in a general method chapter, in addition to the primary validation, the respective USP chapter (e.g., USP <73>) would also apply3.
Charles River Laboratories has successfully satisfied all criteria as outlined in USP <1223>, therefore, users can leverage the pre-existing primary validation data and concentrate on satisfying the test requirements as outlined in USP <73>. The primary validation also includes using several slow-growing organisms and molds relevant to the CGT manufacturing environment. Although the primary validation can be utilised, users must refer to USP <73> and always include relevant environmental isolates in method suitability testing of any RMM system8.
The updated publication of USP <1071> and USP <73>, effective August 2025, significantly impacts end users by reducing the validation burden for successfully implementing RMMs included in the general chapters. Manufacturers of compounded sterile preparations, nuclear medicines, and ATMPs have a simple and effective blueprint to include ATP-bioluminescence in their microbiological control strategies4.
References
- https://www.gmp-compliance.org/gmp-news/european-pharmacopoeia-chapter-2-6-27-microbiological-examination-of-cell-based-preparations-revised#:~:text=A%20whole%20series%20of%20drafts,2
- Eur. 2.6.27 Microbiological Examination of cell-based Preparations
- USP <1071>: Rapid Microbial Test for Release of Sterile Short-Life Products: A Risk-Based Approach
- Risk Assessment Approach to Microbiological Controls of Cell Therapies, PDA Journal
- Demonstration of ATP-Bioluminescence for Detection of Microbial Contaminants in T-Cells Using Celsis Adapt™ – Charles River Laboratories
- Protocol No. 21-074, Celsis Adapt™ Cell Equivalency Fact Sheet
- USP <1223>: Validation of Alternative Microbiological Methods
- USP <73>: ATP Bioluminescence-Based Microbiological Methods for the Detection of Contamination in Short-Life Products
About the author:
Dr Jeff Domsic is currently a Product Manager for the Celsis brand of quality control microbiology test products and services at Charles River Laboratories, and has over 19 years of industry experience. He earned his PhD in Molecular Biology from the University of Iowa.